C

HAPTER

14

T

ROUBLESHOOTING

14–4

www.zoll.com

9650-0904-01 Rev. K

Recorder

Pacer

Symptom

Recommended Action

1. CHECK RECORDER message

Recorder out of paper.

Remove paper, check paper type, check recorder for
paper jam, and reload paper.

Recorder door is open.

Paper is loaded upside down.

2. Recorder makes stuttering sound

when activated.

Check recorder for paper jam.

3. Light or poor quality printing

Ensure correct paper is in use.

Ensure paper is installed grid side against recorder print
head.

Recorder print head requires cleaning.

4. Summary Report will not print

when Report/Print Chart softkeys
are activated.

15 seconds have not elapsed since one of the events
that trigger Summary Report to record have occurred.
Wait 15 seconds, then try again.

Symptom

Recommended Action

1. CHECK PADS message

Ensure therapy electrodes are connected to the
OneStep cable.

Ensure electrode gel is not dry. Replace therapy
electrodes if necessary.

Ensure good electrode-to-patient contact.

Check integrity of OneStep cable by plugging into test
connector. CHECK PADS should disappear.

2. No stimulus marker (

) is

present on the ECG trace.

Ensure unit is in PACER mode.

Ensure PACER RATE (ppm) is set greater than patient
heart rate.

3. No ventricular capture beat

appears after stimulus marker on
ECG display.

Check patient’s pulse.

Increase output current.

Ensure therapy electrodes are making good contact
with the patient.

Select different ECG Lead configuration.

Review therapy electrode placement.