ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 167

Advertising
background image

Defibrillator Specifications

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

A–5

Heart rate accuracy

±5%

Heart rate alarm

Screen icon indicates activated/deactivated status. User selectable.

Tachycardia: 60 to 280 beats per minute
Bradycardia: 20 to 100 beats per minute

Tall T-wave Rejection

< 0.8 mV

Heart Rate Averaging

The R Series averages the interval between the last 5 detected beats.
On startup, the R Series averages the rate between detected beats
once two beats are detected, until a full 5 beats have been received.
The rate is updated every beat. After this condition is met, the meter is
updated every beat with an average of the last 5 beats.

If a period of time greater than 5 seconds elapses without a beat
detected, the heart rate meter reports a rate of 0 bpm, which is
repeated every 5 seconds.

Accuracy and Response
Time to Irregular Rhythm

Averaging over 5 R -R intervals, per AAMI EC 13:2002:

Ventricular bigeminy (Figure 3a) - 80.5 bpm

Slow alternating ventricular bigeminy (Figure 3b) - 60.5 bpm

Slow alternating ventricular bigeminy (Figure 3c) - 120.5 bpm

Bidirectional systoles (Figure 3d) - 93.3 bpm

Response Time to
Change in Heart Rate

Average response time: 3.37 seconds

Time to Alarm for
Tachycardia

Minimum time to alarm: 3.66 seconds
Maximum time to alarm: 8.41 seconds

Leads Off Sensing

A dc current of 0.04 uA per lead wire is supplied to the patient.

Active Noise
Suppression

The sum of all leadwire currents is returned via the active noise
suppression leadwire:

0.08 uA DC in 3 lead mode

0.16 uA DC in 5 lead mode

Pacemaker option

Type

VVI demand; asynchronous (fixed rate) when used without ECG leads
or in asynchronous (Async) pacing mode

Pulse type

Rectilinear, constant current

Pulse duration

40 ms ±2 ms

Pulse amplitude (output)

Variable 0 mA to 140 mA ±5% or 5 mA, whichever is greater
Increments/decrements by a value of 2 mA

Pacing rate

30 to 180 pulses per minute (ppm) ±1.5%
Increments/decrements by a value of 2 ppm

Output protection

Fully defibrillator-protected and isolated

Recorder and Stripchart Printer

Paper

80 mm thermal (grid width)
90 mm (paper width)

Speed

25 mm/s

Delay

6 seconds

Annotations

Time, date, defibrillation energy, heart rate, pacer output, QRS
synchronization marker, ECG size, ECG lead, alarm, defibrillator test
result, analyze ECG, diagnostic bandwidth

Messages: ANALYSIS HALTED, NOISY ECG, SHOCK ADVISED,
NO SHOCK ADV.