of the unit. To switch between paddles and hands-free therapy electrodes, remove the OneStep
cable from the apex paddle and connect the hands-free therapy electrodes to the cable.

The Advisory function cannot be activated unless hands-free therapy electrodes are attached to
the OneStep cable and used as the ECG monitoring lead.

The R Series can monitor the patient’s ECG while pacing without the need for a separate ECG
cable and ECG electrodes. This also allows demand pacing when separate ECG electrodes are
either not connected, or unavailable. OneStep pacing capability requires the OneStep Pacing
cable along with OneStep Pacing electrodes, or OneStep Complete electrodes.

Note:

The ZOLL OneStep electrodes, MFE Pads, Pediatric MFE Pads, stat-padz

®

, and ECG

electrodes are disposable, single-use items.

You should always check the expiration date on the electrode packaging. Do not use expired
electrodes, which might result in false patient impedance readings and affect the level of
delivered energy, or cause burns.

The R Series defibrillator reads and reports the expiration date for all OneStep electrodes
(except for OneStep Basic). When these electrodes exceed their expiration date, the Code
Readiness indicator will change to a red “X.”

Note:

ZOLL electrodes contain no hazardous materials and may be disposed of in general
trash unless contaminated with pathogens. Use appropriate precautions when
disposing of contaminated electrodes.

When the patient is less than 8 years old or weighs less than 55 lb. (25 kg), use ZOLL OneStep
pediatric defibrillation electrodes. Do not delay therapy to determine the patient’s exact age or
weight.

Batteries

R Series products use an easily replaced rechargeable lithium-ion battery pack (the ZOLL
SurePower battery pack). A new, fully charged battery pack typically delivers more than 5
hours of ECG monitoring. Use of other functions (such as the defibrillator, printer, or
pacemaker) reduces this time.

When a LOW BATTERY message appears on the display and the unit emits two beeps in
conjunction with the displayed message, the battery must be replaced and recharged.

You can charge the battery by either of the following methods:

Internal charging — plug the R Series into an AC power supply to automatically begin
charging the installed battery pack. The front panel battery indicator operates as follows:

This symbol on the electrode package is accompanied by the expiration date.

When the indicator is:

It means:

Steady yellow

Battery is charging

Steady green

Battery is charged