150j sel, No shock adv, 3 press shock – ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 54: Charging 76j, Shock advised

Advertising
background image

C

HAPTER

3

A

UTOMATED

E

XTERNAL

D

EFIBRILLATOR

(AED) O

PERATION

3–6

www.zoll.com

9650-0904-01 Rev. K

When a nonshockable rhythm is detected, the unit displays a NO SHOCK ADV. message.
Immediately begin chest compressions and continue other treatment per protocol.

If the patient’s rhythm is shockable, the unit displays a SHOCK ADVISED message. The
defibrillator begins charging automatically to the preconfigured energy setting and displays a
CHARGING message.

When charging is completed, the monitor displays the energy level to which the defibrillator
has charged, XXXJ READY.

3 Press SHOCK

WARNING!

Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.

Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient’s
body to come in contact with metal objects, such as a bed frame, as unwanted pathways
for defibrillation current may result.

150J SEL.

ECG

NO SHOCK ADV.

Manual

Mode

Analyze

NO
SHOCK
ADVISED

CHARGING 76J

Manual

Mode

SHOCK ADVISED

ECG

Analyze