ZOLL R Series Monitor Defibrillator Rev K Operators Guide User Manual

Page 157

Advertising
background image

Monitor

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

14–3

6. Noisy ECG, artifact, wandering

baseline

Consider 1 – 21Hz filter bandwidth (see R Series
Configuration Manual
).

Prepare the patient’s skin prior to electrode attachment.

Check for proper adhesion of electrodes to patient.

Reduce or eliminate ECG artifact due to electrode or
patient cable movement. Route cables so that they don’t
pull on electrodes or swing excessively.

Ensure patient is motionless.

Check for possible excessive radio frequency
interference.

7. Poor ECG signal amplitude,

calibration pulse normal

Select another lead.

Apply new electrodes using different placement.

8. Inconsistent QRS beep or heart

rate

Select another lead.

Alter ECG electrode placement.

9. Sync marker is absent or

inconsistent with QRS waveform
on display and recorder

Ensure device is in Sync mode.

Ensure that device in NOT in Remote Sync mode.

Change ECG lead selection.

Alter ECG electrode placement.

Paper too narrow. It should be 90 mm wide.

10. No ECG trace or dashed line on

display

Device is in Remote Sync mode. Press the
Sync On/Off softkey to exit Remote Sync mode.

11. CPR Remaining Timer is enabled

but fails to decrement

Turn unit off for 10 seconds and then turn it back on
again.

12. Alarm tone sounds when unit is

turned on but alarm feature is not
enabled

Press the ALARM SUSPEND button.

Symptom

Recommended Action