A8-18 – Yokogawa EJX930A User Manual

Page 168

Advertising
background image

<Appendix 8. Advanced Diagnostics>

A8-18

IM 01C25R03-01E

Checklist (2/5)

No.

Items

Parameters

Result

Example

8 Alarm status

• Check the alarm status shown in

DIAG_ERR.

• Check that the alarm status of

“Outside Diagnosis Range” is not

shown in DIAG_ERR.

DIAG_ERR (Index: 2086)

A Blocking

Large Fluctuation of Low Side

Large Fluctuation of High Side

Low Side Blocking

High Side Blocking

B Blocking

Invalid Ref BlkF

Invalid Ref fSPh

Invalid Ref fSPl

Invalid Ref fDP

Outside Diagnosis Range

Reflect Blockage to PV/SV/TV Status 

9 ILBD parameters

• Record the values of parameters for

ILBD operation.

• Check the status of parameters for

ILBD operation.

*: Record the value after checked

that the status of each parameter is

“GOOD”.

DIAG_PERIOD (Index: 2084)

180

DIAG_LIM (Index: 2094)

DIAG_LIM [1]

3.000000

DIAG_LIM [2]

0.300000

DIAG_LIM [3]

5.000000

DIAG_LIM [4]

0.500000

DIAG_LIM [5]

5.000000

DIAG_LIM [6]

0.500000

DIAG_LIM [7]

0.600000

DIAG_LIM [8]

-0.600000

DIAG_LIM [9]

1.000000

DIAG_LIM [10]

0.050000

DIAG_COUNT (Index: 2095)

3

REFERENCE_TIME (Index: 2096)

16:22:55.876

01/31/2008

REFERENCE_FDP* (Index: 2097)

7.43245e-9

REFERENCE_FSPL* (Index: 2098)

7.25765e-9

REFERENCE_FSPH* (Index: 2099)

7.18374e-9

REFERENCE_DPAVG* (Index: 2101)

5.364248

VALUE_TIME (Index: 2102)

16:22:55.876

01/31/2008

VALUE_FDP (Index: 2103) *

7.48562e-9

VALUE_FSPL (Index: 2104) *

7.23277e-9

VALUE_FSPH (Index: 2105) *

7.14085e-9

VALUE_BLKF (Index: 2106) *

-0.287259

VALUE_DPAVG (Index: 2107) *

0.055957

Advertising