Yokogawa EJX930A User Manual

Page 74

Advertising
background image

<9. Parameter Lists>

9-10

IM 01C25R03-01E

Relative

Index Index

Parameter Name

Factory Default

Write

Mode

Explanation

24

2524 UNIT_SEL2

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL2 will be displayed when

"Auto" is selected; user-specified unit at DISPLAY UNIT2 will be

displayed when "Custom" is selected.

25

2525 DISPLAY_UNIT2

Null

O/S

User specified unit to be displayed on display2, which will be

available when "Custom" is selected at UNIT SEL2.

26

2526 EXP_MODE2

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.

27

2527 BLOCK_TAG3

2029

(SECONDARY_

VALUE)

Block tag which includes a parameter to be displayed on

display3.

28

2528 PARAMETER_SEL3

1 (SECONDARY_

VALUE)

O/S

Selection of a parameter to be displayed on display3.

Select a parameter from Table 6.1

29

2529 DISPLAY_TAG3

Null

O/S

Name of block tag to be displayed on display3; up to six

alphanumeric plus a slash [/] and a period [.]

30

2530 UNIT_SEL3

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL3 will be displayed when

"Auto" is selected; user-specified unit at DISPLAY UNIT3 will be

displayed when "Custom" is selected.

31

2531 DISPLAY_UNIT3

Null

O/S

User specified unit to be displayed on display3, which will be

available when "Custom" is selected at UNIT SEL3.

32

2532 EXP_MODE3

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.

33

2533 BLOCK_TAG4

2047 (CAP_

TEMP_VAL)

Block tag which includes a parameter to be displayed on

display4.

34

2534 PARAMETER_SEL4

3 (CAP_TEMP_

VAL)

O/S

Selection of a parameter to be displayed on display4.

Select a parameter from Table 6.1

35

2535 DISPLAY_TAG4

Null

O/S

Name of block tag to be displayed on display4; up to six

alphanumeric plus a slash [/] and a period [.]

36

2536 UNIT_SEL4

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL4 will be displayed when

"Auto" is selected; user-specified unit at DISPLAY UNIT4 will be

displayed when "Custom" is selected.

37

2537 DISPLAY_UNIT4

Null

O/S

User specified unit to be displayed on display4, which will be

available when "Custom" is selected at UNIT SEL4.

38

2538 EXP_MODE4

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.

39

2539 BAR_GRAPH_

SELECT

1 (display)

O/S

Selection of bar graph indicator.

40

2540 DISPLAY_CYCLE

3 (2.4s)

O/S

Duration of display cycle. (Time unit: 1=800ms)

41

2541 TEST40

0

Not used for EJX.

42

2542 BLOCK_TAG5

Null

Block tag which includes a parameter to be displayed on

display5.

43

2543 PARAMETER_SEL5

4008

O/S

Selection of a parameter to be displayed on display5.

Select a parameter from Table 6.1

44

2544 DISPLAY_TAG5

Null

O/S

Name of block tag to be displayed on display5; up to six

alphanumeric plus a slash [/] and a period [.]

45

2545 UNIT_SEL5

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL5 will be displayed when

"Auto" is selected; user-specified unit at DISPLAY UNIT5 will be

displayed when "Custom" is selected.

46

2546 DISPLAY_UNIT5

Null

O/S

User specified unit to be displayed on display5, which will be

available when "Custom" is selected at UNIT SEL5.

47

2547 EXP_MODE5

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.

48

2548 BLOCK_TAG6

Null

Block tag which includes a parameter to be displayed on

display6.

49

2549 PARAMETER_SEL6

4108

O/S

Selection of a parameter to be displayed on display6.

Select a parameter from Table 6.1

50

2550 DISPLAY_TAG6

Null

O/S

Name of block tag to be displayed on display6; up to six

alphanumeric plus a slash [/] and a period [.]

51

2551 UNIT_SEL6

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL6 will be displayed when

"Auto" is selected; user-specified unit at DISPLAY UNIT6 will be

displayed when "Custom" is selected.

Advertising