2 optional specifications, Optional specifications -2 – Yokogawa EJX930A User Manual

Page 83

Advertising
background image

<10. General Specifications>

10-2

IM 01C25R03-01E

10.2 Optional Specifications

For items other than those described below, refer to each User’s Manual.

Item

Description

Code

PID function

PID control function

LC1

Data configuration at factory

Software damping

CC

Software Download Function

Based on Fieldbus Foundation Specification (FF-883)

Download class: Class1

EE

Advanced diagnostics

*1

Multi-sensing process monitoring

• Impulse line blockage detection

*2

• Heat trace monitoring

DG1

*1: This option code must be specified with option code

EE.

*2: The change of pressure fluctuation is monitored and then detects the impulse line blockage. See TI 01C25A31-01E for detailed

technical information required for using this function.

10.3 Optional Specifications (For Explosion Protected type)

Item

Description

Code

Factory

Mutual (FM)

FM Explosionproof

*2

Applicable Standard: FM3600, FM3615, FM3810, ANSI/NEMA 250

Explosionproof for Class I, Division 1, Groups B, C and D, Dust-ignitionproof for Class II/III,

Division 1, Groups E, F and G, in Hazardous locations, indoors and outdoors (NEMA 4X)

Temperature class: T6, Amb. Temp.: –40 to 60°C (–40 to 140°F)

FF1

FM Intrinsically Safe and Nonincendive

*2

Applicable Standard: FM3600, FM3610, FM3611, FM3810, ANSI/NEMA 250, IEC60079-27

Intrinsically Safe for Class I,II, & III, Division 1, Groups A,B,C,D,F & G, Entity, FISCO.

Class I, Zone 0, AEx ia IIC,

Enclosure: "NEMA 4X", Temp. Class: T4, Amb. Temp.: -40 to 60°C (-40 to 140°F).

Intrinsically Apparatus Parameters :

[FISCO (IIC)] Ui=17.5 V, Ii=380 mA, Pi=5.32 W, Ci=3.52 nF, Li=0 µH

[FISCO (IIB)] Ui=17.5 V, Ii=460 mA, Pi=5.32 W, Ci=3.52 nF, Li=0 µH

[Entity] Ui=24 V, Ii=250 mA, Pi=1.2 W, Ci=3.52 nF, Li=0 µH

Sensor Circuit: Uo=6.51 V, Io=4 mA, Po=6 mW, Co=34 µF, Lo=500 mH

Nonincendive for Class I, Division 2, Groups A, B, C and D, NIFW, FNICO

Class I, Zone 2, Group IIC, NIFW, FNICO

Class II, Division 2, Groups F&G, and Class III, Division 1

Enclosure: "NEMA 4X", Temp. Class: T4, Amb. Temp.: -40 to 60°C (-40 to 140°F)

Nonincendive Apparatus Parameters: Vmax.= 32 V, Ci = 1.76 nF, Li = 0 µH

FS15

ATEX

ATEX Flameproof

*2

Applicable Standard: EN 60079-0, EN 60079-1, EN 60079-31

Certificate: KEMA 07ATEX0109 X

II 2G, 2D Ex d IIC T6...T4 Gb, Ex tb IIIC T85°C Db IP6X

Degree of protection: IP66/IP67

Amb. Temp. (Tamb) for gas-proof :

T4; –50 to 75°C (–58 to 167°F), T5; –50 to 80°C (–58 to 176°F), T6; –50 to 75°C (–58 to

167°F)

Max. process Temp. for gas-proof (Tp): T4; 120°C (248°F), T5; 100°C (212°F), T6; 85°C

(185°F)

Max. surface Temp. for dust-proof: T85°C (Tamb: –30 to 75°C, Tp: 85°C)

*1

KF22

ATEX Intrinsically Safe

*2

Applicable Standards: EN 60079-0:2009, EN 60079-11:2012, EN 60079-26:2007

Certificate: KEMA 06ATEX0278X

II 1G, 2D Ex ia IIC/IIB T4 Ga Ex ia IIIC T85°C T100°C T120°C Db Type of protection: IP66/IP67

Amb. Temp. for EPL Ga: –40 to 60°C (–40 to 140°F)

Amb. Temp. for EPL Db: –30 to 60°C

*1

Max. Process Temp. (Tp): 120°C (248°F)

Max. Surface Temp. for EPL Db. T85°C (Tp:80°C), T100°C (Tp:100°C), T120°C (Tp:120°C)

Ambient Humidity: 0 to 100% (No condensation)

Electrical data:

Supply/Output circuit (terminals + and –)

[FISCO (IIC)] Ui=17.5 V, Ii=380 mA, Pi=5.32 W, Ci=3.52 nF, Li=0 μH

[FISCO (IIB)] Ui=17.5 V, Ii=460 mA, Pi=5.32 W, Ci=3.52 nF, Li=0 μH

[Entity] Ui=24 V, Ii=250 mA, Pi=1.2 W, Ci=3.52 nF, Li=0 μH

External Temperature Input circuit (connector)

Uo=7.63 V, Io=3.85 mA, Po=0.008 W, Co=4.8 μF, Lo=100 mH

KS26

Advertising
This manual is related to the following products: