Yokogawa EJX930A User Manual

Page 60

Advertising
background image

<8. Device Information>

8-3

IM 01C25R03-01E

Table 8.5

Contents of DEVICE_STATUS_5

(index 1049)

Hexadecimal Display through

DD

Description

0x80000000

Diff Pressure Span

Trim Error (AL-50)

Trimming range error for

differential pressure span

0x40000000

Diff Pressure Zero

Trim Error (AL-50)

Trimming range error for

differential pressure zero

0x20000000

Static Pressure

Span Trim Error

(AL-51)

Trimming range error for

static pressure span

0x10000000

Static Pressure

Zero Trim Error

(AL-51)

Trimming range error for

static pressure zero

0x08000000

External Temp

Span Trim Error

(AL-52)

Trimming range error

for external temperature

span

0x04000000

External Temp Zero

Trim Error (AL-52)

Trimming range error

for external temperature

zero

Table 8.6

Contents of DEVICE_STATUS_6

(index 1050)

Hexadecimal Display through

DD

Description

0x80000000

PID in O/S mode

(AL-70)

PID block is in O/S mode

0x40000000

PID in MAN mode

(AL-70)

PID block is in MAN

mode

0x20000000

PID Non-Schduled

(AL-70)

PID block is not

Scheduled

0x10000000

SC in O/S mode

(AL-71)

SC block is in O/S mode

0x08000000

SC in MAN mode

(AL-71)

SC block is in MAN mode

0x04000000

SC Non-Schduled

(AL-71)

SC block is not

scheduled

0x02000000

IT in O/S mode

(AL-72)

IT block is in O/S mode

0x01000000

IT in MAN mode

(AL-72)

IT block is in MAN mode

0x00800000

IT Non-Schduled

(AL-72)

IT block is not scheduled

0x00400000

IS in O/S mode

(AL-73)

IS block is in O/S mode

0x00200000

IS in MAN mode

(AL-73)

IS block is in MAN mode

0x00100000

IS Non-Schduled

(AL-73)

IS block is not scheduled

0x00080000

AR in O/S mode

(AL-74)

AR block is in O/S mode

0x00040000

AR in MAN mode

(AL-74)

AR block is in MAN mode

0x00020000

AR Non-Schduled

(AL-74)

AR block is not

scheduled

0x00000002

LCD Display

Outside Range

Limit (AL-79)

Displayed value exceeds

limit

0x00000001

LCD Display Config

Error (AL-79)

Specified settings for

Display do not meet the

conditions.

Table 8.7

Contents of DEVICE_STATUS_7

(index 1051)

Hexadecimal

Display through

DD

Description

0x80000000

Sensor TB

Simulation Active

(AL-80)

Simulation function of

Sensor TB is active

0x40000000

Flow TB

Configuration Error

(AL-81)

Configuration error

occurs in Flow TB

0x20000000

Flow TB Algorithm

Error 1 (AL-82)

Configuration error

0x10000000

Flow TB Algorithm

Error 2 (AL-82)

Configuration error

0x08000000

Flow TB Algorithm

Error 3 (AL-82)

Configuration error

0x04000000

Flow TB Algorithm

Error 4 (AL-82)

Configuration error

0x02000000

Flow TB Algorithm

Error 5 (AL-82)

Configuration error

0x01000000

Flow TB Algorithm

Error 6 (AL-82)

Configuration error

0x00800000

Flow TB Check

Sum Error (AL-83)

Precise flow setting

CHECK SUM error

0x000040000* Flg Temp High

Alarm (AL-87)

Flange Temperature

High Alarm occurs.

0x000002000* Flg Temp Low

Alarm (AL-87)

Flange Temperature

Low Alarm occurs.

0x000001000* Outside Diagonosis

Range (AL-89)

VALUE_DPAVG is

outside range.

0x000000800* Invalid Ref fDP

(AL-88)

VALUE_FDP under

normal conditions is

not up to the reference

fluctuation level required

to blockage detection.

0x000000400* Invalid Ref fSPl

(AL-88)

VALUE_FSPL under

normal conditions is

not up to the reference

fluctuation level required

to blockage detection.

0x000000200* Invalid Ref fSPh

(AL-88)

VALUE_FSPH under

normal conditions is

not up to the reference

fluctuation level required

to blockage detection.

0x000000100* Invalid Ref BlkF

(AL-88)

VALUE_BLKF under

normal conditions is

not up to the reference

fluctuation level required

to blockage detection.

0x000000080* B Blocking (AL-89) B Blocking (both- side

blockage) is detected.

0x000000040* High Side Blocking

(AL-89)

High-pressure-side

blockage is detected.

0x000000020* Low Side Blocking

(AL-89)

Low-pressure-side

blockage is detected.

0x000000010* Large Fluctuation of

High Side (AL-89)

Pressure fluctuation

amplitude of high-

pressure side is large.

0x000000008* Large Fluctuation

of Low Side (AL-89)

Pressure fluctuation

amplitude of low-

pressure side is large.

0x000000004* A Blocking (AL-89) A Blocking (single-side

blockage) is detected.

*: option code /DG1