<9. Parameter Lists>

9-11

IM 01C25R03-01E

Relative

Index Index

Parameter Name

Factory Default

Write

Mode

Explanation

52

2552 DISPLAY_UNIT6

Null

O/S

User specified unit to be displayed on display6, which will be

available when "Custom" is selected at UNIT SEL6.

53

2553 EXP_MODE6

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.

54

2554 BLOCK_TAG7

Null

Block tag which includes a parameter to be displayed on

display7.

55

2555 PARAMETER_SEL7

4208

O/S

Selection of a parameter to be displayed on display7.

Select a parameter from Table 6.1

56

2556 DISPLAY_TAG7

Null

O/S

Name of block tag to be displayed on display7; up to six

alphanumeric plus a slash [/] and a period [.]

57

2557 UNIT_SEL7

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL7 will be displayed when

"Auto" is selected; user-specified unit at DISPLAY UNIT7 will be

displayed when "Custom" is selected.

58

2558 DISPLAY_UNIT7

Null

O/S

User specified unit to be displayed on display7, which will be

available when "Custom" is selected at UNIT SEL7.

59

2559 EXP_MODE7

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.

60

2560 BLOCK_TAG8

Null

Block tag which includes a parameter to be displayed on

display8.

61

2561 PARAMETER_SEL8

2035

O/S

Selection of a parameter to be displayed on display8.

Select a parameter from Table 6.1

62

2562 DISPLAY_TAG8

Null

O/S

Name of block tag to be displayed on display8; up to six

alphanumeric plus a slash [/] and a period [.]

63

2563 UNIT_SEL8

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL8 will be displayed when

"Auto" is selected; user-specified unit at DISPLAY UNIT8 will be

displayed when "Custom" is selected.

64

2564 DISPLAY_UNIT8

Null

O/S

User specified unit to be displayed on display8, which will be

available when "Custom" is selected at UNIT SEL8.

65

2565 EXP_MODE8

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.

66

2566 BLOCK_TAG9

Null

Block tag which includes a parameter to be displayed on

display9.

67

2567 PARAMETER_SEL9

2064

O/S

Selection of a parameter to be displayed on display9.

Select a parameter from Table 6.1

68

2568 DISPLAY_TAG9

Null

O/S

Name of block tag to be displayed on display9; up to six

alphanumeric plus a slash [/] and a period [.]

69

2569 UNIT_SEL9

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL9 will be displayed when

"Auto" is selected; user-specified unit at DISPLAY UNIT9 will be

displayed when "Custom" is selected.

70

2570 DISPLAY_UNIT9

Null

O/S

User specified unit to be displayed on display9, which will be

available when "Custom" is selected at UNIT SEL9.

71

2571 EXP_MODE9

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.

72

2572 BLOCK_TAG10

Null

Block tag which includes a parameter to be displayed on

display10.

73

2573 PARAMETER_SEL10 2314

O/S

Selection of a parameter to be displayed on display10.

Select a parameter from Table 6.1

74

2574 DISPLAY_TAG10

Null

O/S

Name of block tag to be displayed on display10; up to six

alphanumeric plus a slash [/] and a period [.]

75

2575 UNIT_SEL10

0 (Auto)

O/S

Selection of unit to be displayed. The unit of the parameter

which is selected at PARAMETER SEL10 will be displayed

when "Auto" is selected; user-specified unit at DISPLAY UNIT10

will be displayed when "Custom" is selected.

76

2576 DISPLAY_UNIT10

Null

O/S

User specified unit to be displayed on display10, which will be

available when "Custom" is selected at UNIT SEL10.

77

2577 EXP_MODE10

0

O/S

Selection of the displayed value in exponent such as x1, x10,

x100, and x1000.