93 • Replacing a Sensor

10,000 ppm Gas Tracer Operator’s Manual

dummy sensor installed.

7. Carefully insert the replacement sensor in the correct socket.

Figure 21: Replacing a Sensor

CAUTION:

When replacing the ppm sensor, verify that the sensor is
properly aligned with its socket before inserting it into the
socket. The ppm combustible sensor has two non-polarized
contacts which must line up with the contacts in its socket.
Forcing the sensor into its socket may damage the sensor or the
socket.

8. Be sure that the dummy sensors are installed correctly. Make sure

the flat side is up and the hollow side is down.

9. Place the sensor gasket back into the chamber ensuring that it seals

with the sensors and the edge of the chamber.

10.Place the flow chamber in position over the sensor area and press it

into the case until it is flush with the back of the case.

11.Tighten the top screw of the flow chamber completely.

PPM

Combustible

Sensor

Flow

Chamber

HC

PPM

O2

HC

LEL

Sensor
Gasket

Knob

HC

VOL

CO

Exhaust