Password, Performing a low flow setpoint adjustment – RKI Instruments Gas Tracer 10,000 ppm User Manual

Page 87

Advertising
background image

10,000 ppm Gas Tracer Operator’s Manual

Using Maintenance Mode • 80

NOTE:

A password is always required to enter Maintenance Mode
even if PASSWORD is turned OFF. You may change the
factory set password by turning PASSWORD ON and
entering a new password when prompted for it.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the PASSWORD

menu item. Press and release the POWER ENTER button. The
screen indicates the current setting.

2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

3. If you selected OFF, press and release the POWER ENTER button

to save the setting and return to the main menu.

If you selected ON, press and release the POWER ENTER button.
ON

is replaced by the four characters making up the current

password.

4. Use the AIR▲ or (SHIFT)▼ button to select a number from 0 to 9.

5. Press and release the POWER ENTER button to enter the selection

and advance to the next character.

6. Repeat steps 4 and 5 to select the remaining characters. When you

press and release the POWER ENTER button to enter the last
number, the password is saved and you return to the main menu.

Performing a Low Flow Setpoint Adjustment

If your instrument repeatedly goes into low flow alarm and you have
tried all of the recommendations in the Troubleshooting section, you
may need to perform a low flow setpoint adjustment.

On


MAINTENANCE

PASSWORD

2102


MAINTENANCE

PASSWORD