10,000 ppm Gas Tracer Operator’s Manual

Calibrating and Bump Testing the PPM Com-

To return to the BUMP screen in Calibration Mode, press
POWER ENTER.

If the sensor fails the bump test, see “Troubleshooting” on page 82
to investigate the cause of the failure and replace the failed sensor
if necessary.

Disconnect the humidifier from the probe.

NOTE:

Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.

8. When AUTO CAL in the BUMP--SET menu item in Maintenance

Mode is set to ON:

If the ppm combustible sensor passes the bump test,

The following screen appears:

To view the bump test gas reading press the AIR▲ or (SHIFT)▼
button.

You can scroll between the bump test results and the

bump test gas reading with the AIR▲ or (SHIFT)▼ buttons.


MAINTENANCE

BUMP

P


ppm

MAINTENANCE

CH

4

BUMP

1000


ppm

MAINTENANCE

CH

4

BUMP