10,000 ppm Gas Tracer Operator’s Manual

Calibration Mode • 40

2. Press and release the POWER ENTER button to enter the AIR CAL

menu item.

You will be prompted to choose between WET and DRY.

3. Use the AIR▲ or (SHIFT)▼ buttons to display the DRY setting and

press and release the POWER ENTER button. The unit will display
the current fresh air reading and prompt you to push the AIR▲
button. Do not press the AIRbutton at this time.

4. Install the demand flow regulator onto the zero air cylinder.

5. Connect the sample tubing to the demand flow regulator.

6. Install the hose and probe to the Gas Tracer.

7. Connect the tubing from the demand flow regulator to the rigid

tube on the probe. Allow the zero air to flow for two minutes.

8. Hold down the AIR▲ button. The unit will prompt you to continue

holding the AIR▲ button.

9. When the unit prompts you to release the AIR▲ button, release it.

The instrument will perform a fresh air adjustment on the ppm


MAINTENANCE

WET

50


ppm

MAINTENANCE

CH

4

PUSH AIR

AIr
CAL

MAINTENANCE

HOLD AIR