83 • Troubleshooting

10,000 ppm Gas Tracer Operator’s Manual

• Display indicates

“FAIL SYSTEM”
during startup
along with an error
code.

• A circuit fault has

occurred on one
of the PCB’s.

1. If the error code is 031,

indicating a FLASH memory
failure, press and release the
RESET SILENCE button to
continue to measuring mode
keeping in mind that there will
be no datalogging function.

2. If the error code is anything

else, the instrument cannot be
used.

3. Contact RKI Instruments, Inc.

for further instruction if either
1 or 2 occur.

• Display indicates

“FAIL CLOCK”
during startup.

• An internal clock

failure has
occurred
because the date
in the unit’s
memory is not a
valid date.

1. Enter normal operation by

pressing and releasing the
RESET SILENCE button to
use the unit temporarily.

2. Attempt to set the date in the

DATE menu item in
Maintenance Mode.

3. If you cannot set the date,

contact RKI Instruments, Inc.
for further instruction.

• Heart symbol at

the top of the
screen becomes
steadily on or
disappears

• A microprocessor

error has
occurred.

• Contact RKI Instruments, Inc.

for further instruction.

• The LCD shows

abnormally high or
low readings but
other gas
detection
instruments do
not.

• The Gas Tracer

may need to be
recalibrated.

• The ppm sensor

may need
replacement.

1. Recalibrate the unit.
2. If the difficulties continue,

replace the ppm sensor and
calibrate the ppm channel.

3. If the difficulties continue,

contact RKI Instruments for
further instruction.

Table 5: Troubleshooting the Gas Tracer

Symptoms

Probable

Causes

Recommended Action