10,000 ppm Gas Tracer Operator’s Manual

Calibrating and Bump Testing the PPM Com-

Performing a Bump Test

This section describes how to bump test the PPM combustible gas
sensor.

NOTE:

The BUMP DSP menu item in Maintenance Mode is factory
set to OFF. The BUMP menu item will not appear unless
BUMP DSP

is set to ON. See “Turning the Bump Test

Function On or Off” on page 68 for instructions.

NOTE:

The Gas Tracer uses parameters defined in the BUMP--SET
menu item of Maintenance Mode while performing a bump
test. To view or update these parameters, see “Updating the
Bump Test Parameters” on page 69.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the BUMP menu

item.

2. Fill the gas bag with ppm calibration gas as described in “Filling

the Gas Bag” on page 98.

3. Press and release the POWER ENTER button. You will be

prompted to choose between WET and DRY. Use the WET setting.

NOTE:

The DRY setting is for factory use only. Do not perform a
bump test using the DRY setting.

4. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and

press and release the POWER ENTER button. The bump test gas
value will be displayed.


MAINTENANCE

BUMP

1000


ppm

MAINTENANCE

CH

4

BUMP 30