Bp intvl, Updating the bump test interval, Updating the bump test reminder setting – RKI Instruments Gas Tracer 10,000 ppm User Manual

Page 81

Advertising
background image

10,000 ppm Gas Tracer Operator’s Manual

Using Maintenance Mode • 74

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Bump Test Interval

The BP INTVL setting defines the amount of time between bump
tests. The minimum setting is 0 days and the maximum setting is 30
days. The factory setting is 30 days.

NOTE:

If the BUMP DSP menu item is set to OFF, this menu item
does not appear.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the BP INTVL

menu item. Press and release the POWER ENTER button. The
screen indicates the current setting.

2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

Tip:

Press and hold the AIR▲ or (SHIFT)▼ button to rapidly scroll
through the settings.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Bump Test Reminder Setting

With BP RMNDR set to ON, the Gas Tracer will give an indication at
start up if it is due for bump testing. The type of indication will
depend on the BP EXPRD setting (see below).

With BP RMNDR set to OFF (factory setting), the Gas Tracer will give
no indication at start up if it is due for bump testing.

NOTE:

If the BUMP DSP menu item is set to OFF, this menu item
does not appear.

30


MAINTENANCE

BP INTVL