10,000 ppm Gas Tracer Operator’s Manual

Responding to Alarms • 24

Responding to Alarms

This section describes response to battery, sensor, clock, system, and
microprocessor failure alarms.

Responding to Battery Alarms

WARNING: The Gas Tracer is not operational as a gas monitoring

device during a dead battery alarm. Take the Gas Tracer
to a non-hazardous area and replace or recharge the
batteries as described in “Replacing or Recharging the
Batteries” on page 85.

The Gas Tracer is fully functional during a low battery warning.
However, only a limited amount of operating time remains,
approximately 1 - 2 hours. The amount of time depends on how often
the LCD backlight is used and how often the unit is responding to
alarm conditions. Recharge the battery pack or replace the alkaline
batteries as soon as possible as described in “Replacing or Recharging
the Batteries” on page 85.

Sensor Failure

FAIL appears at the

bottom of the display
and the failed sensor
is indicated

• Alarm LED arrays

flash

Buzzer sounds a double
pulsing tone (two pulses
in quick succession)

Clock Failure

• FAIL CLOCK appears

on the display

• Alarm LED arrays

flash

Buzzer sounds a double
pulsing tone (two pulses
in quick succession)

System Failure

• FAIL SYSTEM

appears on the
display

• Alarm LED arrays

flash

• Buzzer sounds a

double pulsing tone
(two pulses in quick
succession)

• A failure code appears

Microprocessor
Failure

• NOTE: The unit will

not operate if this
alarm occurs.

• Heart indicator is

steadily on or not on
at all

• None

Table 3: Alarm Types and Indications

Alarm Type

Visual Indications

Other Indications