10,000 ppm Gas Tracer Operator’s Manual

Calibration Mode • 44

be replaced by 3 dashes.

7. Press and release the POWER ENTER button. The single

calibration screen displays. The gas reading flashes.

8. Connect the humidifier from the demand flow regulator to the

rigid tube on the probe. Allow the calibration gas to flow for two
minutes.

9. If necessary, use the AIR▲ and (SHIFT)▼ buttons to adjust the

reading to match the concentration listed on the calibration
cylinder.

If the reading already matches the concentration listed on the calibration
cylinder, press and release the AIRS button then press and release the
(SHIFT)T button so that the instrument sees a change and resets the
calibration date.

10.Press and release the POWER ENTER button to save the span

value. The LCD will indicate that the calibration has ended, then
the single calibration menu displays.

11.Disconnect the humidifier from the probe.

NOTE:

Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.

12.Unscrew the demand flow regulator from the calibration cylinder.

13.Use the (SHIFT)▼ button to navigate to the ESCAPE menu item,

then press and release the POWER ENTER button to return to
Measuring Mode.

- - -