Auto cal – RKI Instruments Gas Tracer 10,000 ppm User Manual

Page 66

Advertising
background image

59 • Using Maintenance Mode

10,000 ppm Gas Tracer Operator’s Manual

7. The gas concentration displayed in the calibration value screen

must match the gas concentration listed on the calibration cylinder.
If the concentration matches, go to step 13. If the concentration
does not match, continue with step 8.

8. To adjust the value on the screen, hold down the (SHIFT)▼ button,

then press the DISPLAY(ADJ) button and release both. The
calibration gas value turns into three dashes.

9. Press and release the POWER ENTER button. The sensor’s

calibration gas value will flash.

10.If necessary, use the AIR▲ and (SHIFT)▼ buttons to set the correct

calibration gas value.

11.Press and release the POWER ENTER button to save the change.

12.Use the AIR▲ and (SHIFT)▼ button to scroll to ESCAPE and press

POWER ENTER to return to the calibration gas value screen.

13.With the calibration value screen displayed, press the POWER

ENTER button. AUTO CAL begins to flash and the current gas
reading is displayed. If you want to exit back to the main menu
without completing a calibration, press and release the DISPLAY
(ADJ) button.

14.Connect the humidifier from the demand flow regulator to the

rigid tube on the probe. Allow the gas to flow for two minutes.

15.Press and release the POWER ENTER button to set the calibration

to the programmed values.

If the ppm sensor passed calibration, PASS displays along the

- - -


ppm

MAINTENANCE

CH

4

AUTO CAL

0


ppm

MAINTENANCE

CH

4

AUTO CAL