10,000 ppm Gas Tracer Operator’s Manual

Using Maintenance Mode • 72

Tip:

Press and hold the AIR▲ or (SHIFT)▼ button to rapidly scroll
through the settings.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Calibration Reminder Setting

With CL RMNDR set to ON, the Gas Tracer will give an indication at
start up if it is due for calibration. The type of indication will depend
on the CL EXPRD setting (see below).

With CL RMNDR set to OFF (factory setting), the Gas Tracer will give
no indication at start up if it is due for calibration.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the CL RMNDR

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Calibration Expired Action Setting

This item defines what indication is given during start up when
calibration is due and CL RMNDR is set to ON.

With CL EXPRD set to CONFIRM (factory setting), the Gas Tracer
will give an indication at start up if calibration is due and require the
user to press the RESET SILENCE button to continue.

With CL EXPRD set to NOT USE, if the unit is due for calibration the
Gas Tracer will give an indication at start up that calibration is due
and the unit cannot be used until it is calibrated. With this setting, it is
necessary to calibrate the instrument from Maintenance Mode if
calibration is not performed before it is due. See “Calibrating Using
Auto Calibration” on page 53, “Calibrating Using Single Calibration”
on page 53, or “Calibrating and Bump Testing the PPM Combustible
Sensor” on page 54 for instructions.

On


MAINTENANCE

CL RMNDR