73 • Using Maintenance Mode

10,000 ppm Gas Tracer Operator’s Manual

With CL EXPRD set to NO EFFECT, no confirmation during startup
or calibration is required to use the Gas Tracer if calibration is
overdue.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the CL EXPRD

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIR▲ or (SHIFT) ▼ button to display the desired setting

along the bottom of the screen.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Calibration Check Setting

This item setting defines which sensors are used to update the
calibration reminder screen. C--CHECK can be set to 4-GAS or ALL. If
C--CHECK

is set to 4-GAS, the calibration dates for only the standard

sensors (LEL, O

2

, and CO) are used to determine if calibration is due.

If C--CHECK is set to ALL (factory setting), the calibration dates for
the standard sensors along with the %volume sensor and the ppm
combustible gas sensor are used to determine if a calibration is due.
Since the Gas Tracer only has a ppm sensor, C--CHECK must be set to
ALL.

1. Use the AIRS or (SHIFT)T buttons to display the C--CHECK

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIRS or (SHIFT)T button to display ALL.

CAL

SEL

MAINTENANCE

CONFIRM

ALL


MAINTENANCE

C--CHECK