75 • Using Maintenance Mode

10,000 ppm Gas Tracer Operator’s Manual

1. Use the AIR▲ or (SHIFT)▼ buttons to display the BP RMNDR

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Bump Test Expired Action Setting

This item defines what indication is given during start up when a
bump test is due and BP RMNDR is set to ON.

With BP EXPRD set to CONFIRM (factory setting), the Gas Tracer
will give an indication at start up if a bump test is due and require the
user to press the RESET SILENCE button to continue.

With BP EXPRD set to NOT USE, if the unit is due for a bump test the
Gas Tracer will give an indication at start up that the bump test is due
and the unit cannot be used until it is bump tested. With this setting, it
is necessary to bump test the instrument from the Maintenance Mode
if a bump test is not performed before it is due. See “Performing a
Bump Test” on page 54 and “Calibrating and Bump Testing the PPM
Combustible Sensor” on page 54 for instructions.

With BP EXPRD set to NO EFFECT, no confirmation during startup
or bump test is required to use the Gas Tracer.

NOTE:

If the BUMP DSP menu item is set to OFF, this menu item
does not appear.