Preparing for calibration or bump test, Air cal, Hold air – RKI Instruments Gas Tracer 10,000 ppm User Manual

Page 105: Normal operation fresh air adjustment, Filling the gas bag

Advertising
background image

10,000 ppm Gas Tracer Operator’s Manual

Preparing for Calibration or Bump Test • 98

To calibrate or bump test the Gas Tracer, you will need:

• Known calibrating sample of ppm methane gas.

• A gas collection bag with hose clamp

• A 0.5 LPM fixed-flow regulator or dispensing valve

• Non-absorbent tubing

• 6 inch humidifier tube

Preparing for Calibration or
Bump Test

Perform the following tasks prior to performing a calibration or a
bump test.

Normal Operation Fresh Air Adjustment

Whether you will calibrate the Gas Tracer or perform a bump test, you
will need to set the fresh air readings first.

1. Find a fresh air environment. This is an environment free of toxic

and combustible gases and of normal oxygen content (20.9%).

2. Turn on the unit as described in “Turning On the Gas Tracer” on

page 16. Use the sample hose and probe.

3. With the unit on, perform a fresh air adjust by pressing and

holding the AIR▲ button. The display prompts you to hold the
AIR▲ button.

4. Continue to hold the AIR▲ button until the display prompts you to

release it. The Gas Tracer will set the fresh air reading.

Filling the Gas Bag

Use the following procedure to fill the gas bag with either zero air or
ppm calibration gas.

1. Slide the tubing clamp onto the tubing and connect the tubing onto

the sample bag’s inlet. Leave the clamp unclamped for now.

AIr
CAL

MAINTENANCE

HOLD AIR