10,000 ppm Gas Tracer Operator’s Manual

Start Up • 22

12.The Gas Tracer is now operating in Leak Check Mode.

The combustible gas reading is displayed along with the battery
charge level.

Performing a Fresh Air Adjustment

Before using the Gas Tracer for leak checking, it is recommended to
set the fresh air reading by performing a fresh air adjustment. This
will set the combustible gas channel to zero.

1. Find a fresh-air environment. This is an environment free of toxic

or combustible gases and of normal oxygen content (20.9%).

2. Turn on the unit as described above in “Turning On the Gas Tracer”

on page 16.

3. Press and hold the AIR▲ button. The display will prompt you to

hold the AIR▲ button.

4. Continue to hold the AIR▲ button until the display prompts you to

release the AIR▲ button. The Gas Tracer will set the fresh air
reading for the combustible channel. Once the fresh air adjustment
has finished, start up is complete and the unit is ready for leak
testing.

5. If you would like to calibrate the Gas Tracer while in Leak Check

Mode, follow the same instructions described in “Calibration
Mode” on page 35.