Calibration/bump test supplies and equipment – RKI Instruments Gas Tracer 10,000 ppm User Manual
Page 104
Advertising
97 • Overview
10,000 ppm Gas Tracer Operator’s Manual
Appendix A: Calibrating or Bump Testing with a
Sample Bag
Overview
The Gas Tracer can be calibrated or bump tested with a gas bag
calibration kit instead of a demand flow regulator kit. Appendix A
describes how to use a sample bag calibration kit to calibrate and
bump test the Gas Tracer. A parts list at the end of this appendix lists
spare parts for the calibration kit.
Calibration/Bump Test Supplies and Equipment
Figure 23: Gas Bag Calibration Kit
Calibration Cylinder
Regulator
Gas Bag
Hose Clamp
Calibration Tubing