10,000 ppm Gas Tracer Operator’s Manual

Using Maintenance Mode • 70

increments are 30, 45, 60, and 90 seconds. Once a value is selected,
press and release the POWER ENTER button to save it. The value
will stop blinking.

4. To update the bump test tolerance, use the (SHIFT)▼ button to

scroll to the CHECK value. The factory setting is 30%. This is the
bump test tolerance value and is represented as a percentage. It is
the percentage that the bump test reading can differ from the actual
gas concentration. If the bump test reading differs more, the bump
test will fail. With the cursor next to the check value and CHECK
displayed along the bottom of the screen, press and release
POWER ENTER. The value will begin to flash. Use the AIR▲ or
(SHIFT)▼ buttons to select a new threshold value. The available
values are 10, 20, 30, 40, and 50. Once a value is selected, press and
release the POWER ENTER button to save it. The value will stop
blinking.

5. To update the CAL TIME, use the (SHIFT)▼ button to scroll to the

CAL TIME value. The factory setting is 90 seconds. This is the
length of time the instrument will remain exposed to calibration
gas after a failed bump test if AUTO CAL is set to On. The bump
test time is deducted from the calibration time. For example, if the
CAL TIME is set to 90 seconds and the GAS TIME is set to 30
seconds, if the bump test fails, the Gas Tracer will only be exposed
to gas for an additional 60 seconds. With the cursor next to the cal
time value and CAL TIME displayed along the bottom of the
screen, press and release the POWER ENTER button. The value
will begin to flash. Use the AIR▲ or (SHIFT)▼ buttons to select a
new cal time value. The choices are 90 and 120 seconds. Once a
value is selected, press and release the POWER ENTER button to
save it. The value will stop blinking.

6. To update the AUTO CAL setting, use the (SHIFT)▼ button to

scroll to the AUTO CAL value. The factory setting is ON. With
AUTO CAL set to ON, if a bump test fails, the unit will
automatically begin a calibration. If AUTO CAL is set to OFF and a
bump test fails, nothing will happen. With the cursor next to the
auto cal selection and AUTO CAL displayed along the bottom of
the screen, press and release the POWER ENTER button. The value
will begin to flash. Use the AIR▲ or (SHIFT)▼ buttons to select ON
or OFF. Once a selection is made, press and release the POWER
ENTER button to save it. The value will stop blinking.

7. To exit the BUMP--SET menu, use the (SHIFT)▼ button to scroll to

ESCAPE and press and release the POWER ENTER button.