45 • Calibration Mode

10,000 ppm Gas Tracer Operator’s Manual

Performing a Bump Test

NOTE:

The BUMP DSP menu item in Maintenance Mode is factory
set to OFF. The BUMP menu item will not appear unless
BUMP DSP

is set to ON. See “Turning the Bump Test

Function On or Off” on page 68 for instructions.

NOTE:

The Gas Tracer uses parameters defined in the BUMP--SET
menu item of Maintenance Mode while performing a bump
test. To view or update these parameters, see “Updating the
Bump Test Parameters” on page 69.

1. Install the demand flow regulator onto the calibration cylinder.

2. Connect the humidifier to the demand flow regulator.

3. Install the hose and probe to the Gas Tracer.

4. While in the Calibration Mode menu, use the AIR▲ or (SHIFT)▼

buttons to display the BUMP menu item.

5. Press and release the POWER ENTER button. You will be

prompted to choose between WET and DRY. Use the WET setting.

NOTE:

The DRY setting is for factory use only. Do not perform a
bump test using the DRY setting.

6. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and

press and release the POWER ENTER button. The bump test gas
value screen will be displayed.


MAINTENANCE

BUMP

1000


ppm

MAINTENANCE

CH

4

BUMP 30