One cal – RKI Instruments Gas Tracer 10,000 ppm User Manual

Page 114

Advertising
background image

107 • Calibrating and Bump Testing the PPM Combustible Sensor 10,000 ppm Gas

4. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and

press and release the POWER ENTER button. The gas reading will
be replaced by 3 dashes.

5. Press and release the POWER ENTER button. The single

calibration screen displays for the sensor you selected. The gas
reading flashes.

6. Connect the humidifier from the gas bag to the rigid tube on the

probe. Allow the calibration gas to flow for two minutes.

7. If necessary, use the AIR▲ and (SHIFT)▼ buttons to adjust the

reading to match the concentration listed on the calibration
cylinder.

If the reading already matches the concentration listed on the calibration
cylinder, press and release the AIRS button then press and release the
(SHIFT)T button so that the instrument sees a change and resets the
calibration date.

8. Press and release the POWER ENTER button to save the span

value. The LCD will indicate that the calibration has ended, then
the single calibration menu displays.

9. Disconnect the humidifier from the probe.

NOTE:

Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.

10.Use the (SHIFT)▼ button to navigate to the ESCAPE menu item,

then press and release the POWER ENTER button to return to
Measuring Mode.

- - -


ppm

MAINTENANCE

CH

4

ONE CAL

0


ppm

MAINTENANCE

CH

4

ONE CAL

Advertising