10,000 ppm Gas Tracer Operator’s Manual

Calibration Mode • 38

1. If you did not perform a fresh air adjustment while in Leak Check

Mode, you must perform one before performing an auto
calibration, a single calibration, or a bump test.

2. Press and release the POWER ENTER button to enter the AIR CAL

menu item.

You will be prompted to choose between WET and DRY.

3. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and

press and release the POWER ENTER button. The unit will display
the current fresh air reading and prompt you to push the AIR▲
button.

NOTE:

If you have just performed a DRY fresh air adjustment, allow
the Gas Tracer to sample fresh air for 2 minutes before
continuing.

4. Be sure the Gas Tracer is in a fresh air environment.


MAINTENANCE

AIR CAL


MAINTENANCE

WET

50


ppm

MAINTENANCE

CH

4

PUSH AIR