10,000 ppm Gas Tracer Operator’s Manual

Calibration Mode • 46

7. Before proceeding, confirm that the bump test gas value is the

same as the concentration in the calibration cylinder. If it is not,
adjust the bump test gas value by entering the AUTO CAL menu
item, changing the value there, and reentering the BUMP menu.

8. Connect the humidifier from the demand flow regulator to the

rigid tube on the probe then quickly press and release the POWER
ENTER button. The unit will begin counting down from 30.

9. When AUTO CAL in the BUMP--SET menu item in Maintenance

Mode is set to OFF:

The instrument will display if the ppm combustible sensor
passed or failed with a P (pass) or an F (fail) to the right of the
gas.

To view the bump test gas reading press the AIR▲ or (SHIFT)▼
button. You can scroll between the bump test results and the
bump test gas reading with the AIR▲ or (SHIFT)▼ button.

To return to the BUMP screen, press POWER ENTER.

If the sensor fails the bump test, see “Troubleshooting” on page 82