Beep. set, Cl intvl, Updating the confirmation beep setting – RKI Instruments Gas Tracer 10,000 ppm User Manual
Page 78: Updating the datalog overwrite setting, Updating the calibration interval

71 • Using Maintenance Mode
10,000 ppm Gas Tracer Operator’s Manual
Updating the Confirmation Beep Setting
With BEEP. SET set to ON, the Gas Tracer beeps once every 5 minutes
to verify that it is operating.
With BEEP. SET set to OFF (factory setting), the Gas Tracer does not
sound a confirmation beep.
1. Use the AIR▲ or (SHIFT)▼ buttons to display the BEEP. SET menu
item. Press and release the POWER ENTER button. The screen will
indicate the current setting.
2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Updating the Datalog Overwrite Setting
The OVER WRT menu item does not have any effect on the Gas
Tracer.
Updating the Calibration Interval
This CL INTVL setting defines the amount of time between
calibrations. The minimum setting is 1 day and the maximum setting
is 255 days. The factory setting is 90 days.
1. Use the AIR▲ or (SHIFT)▼ buttons to display the CL INTVL
menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.
2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.
On
MAINTENANCE
BEEP. SET
90
MAINTENANCE
CL INTVL