71 • Using Maintenance Mode

10,000 ppm Gas Tracer Operator’s Manual

Updating the Confirmation Beep Setting

With BEEP. SET set to ON, the Gas Tracer beeps once every 5 minutes
to verify that it is operating.

With BEEP. SET set to OFF (factory setting), the Gas Tracer does not
sound a confirmation beep.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the BEEP. SET menu

item. Press and release the POWER ENTER button. The screen will
indicate the current setting.

2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

3. Press and release the POWER ENTER button to save the setting

and return to the main menu.

Updating the Datalog Overwrite Setting

The OVER WRT menu item does not have any effect on the Gas
Tracer.

Updating the Calibration Interval

This CL INTVL setting defines the amount of time between
calibrations. The minimum setting is 1 day and the maximum setting
is 255 days. The factory setting is 90 days.

1. Use the AIR▲ or (SHIFT)▼ buttons to display the CL INTVL

menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.

2. Use the AIR▲ or (SHIFT)▼ button to display the desired setting.

On


MAINTENANCE

BEEP. SET

90