Test sel, Confirm, B--check – RKI Instruments Gas Tracer 10,000 ppm User Manual
Page 83: Updating the bump test check setting

10,000 ppm Gas Tracer Operator’s Manual
Using Maintenance Mode • 76
1. Use the AIR▲ or (SHIFT)▼ buttons to display the BP EXPRD
menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.
2. Use the AIR▲ or (SHIFT) ▼ button to display the desired setting
along the bottom of the screen.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
Updating the Bump Test Check Setting
This item setting defines which sensors are used to update the bump
test reminder screen. B--CHECK can be set to 4-GAS or ALL. If B--
CHECK
is set to 4-GAS, the calibration dates for only the standard
sensors (LEL, O
2
, and CO) are used to determine if a bump test is due.
If B--CHECK is set to ALL (factory setting), the calibration dates for
the standard sensors along with the %volume sensor and the ppm
combustible gas sensor are used to determine if a bump test is due.
Since the Gas Tracer only has a ppm sensor, B--CHECK must be set to
ALL.
1. Use the AIRS or (SHIFT)T buttons to display the B--CHECK
menu item. Press and release the POWER ENTER button. The
screen will indicate the current setting.
2. Use the AIRS or (SHIFT)T button to display ALL.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
tESt
SEL
MAINTENANCE
CONFIRM
ALL
MAINTENANCE
B--CHECK