47 • Calibration Mode

10,000 ppm Gas Tracer Operator’s Manual

to investigate the cause of the failure and replace the failed sensor
if necessary.

Disconnect the humidifier from the probe.

NOTE:

Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.

10.When AUTO CAL in the BUMP--SET menu item in Maintenance

Mode is set to ON:

If the ppm combustible sensor passes the bump test,

The following screen appears:

To view the bump test gas reading press the AIR▲ or (SHIFT)▼
button. You can scroll between the bump test results and the
bump test gas reading with the AIR▲ or (SHIFT)▼ button.

To return to the BUMP screen, press POWER ENTER.

Disconnect the humidifier from the probe.

P


ppm

MAINTENANCE

CH

4

BUMP

1000


ppm

MAINTENANCE

CH

4

BUMP


MAINTENANCE

BUMP