103 • Calibrating and Bump Testing the PPM Combustible Sensor 10,000 ppm Gas

1. After entering Calibration Mode, the AIR CAL menu item will be

displayed. You must perform a fresh air adjust before performing
an auto calibration, a single calibration, or a bump test.

2. Fill the gas bag with zero air as described in “Filling the Gas Bag”

on page 98.

3. Press and release the POWER ENTER button to enter the AIR CAL

menu item.

You will be prompted to choose between WET and DRY.

4. Use the AIR▲ or (SHIFT)▼ buttons to display the DRY setting and

press and release the POWER ENTER button. The unit will display
the current fresh air reading and prompt you to push the AIR▲
button. Do not press the AIRbutton at this time.

5. Connect the tubing from the gas bag to the rigid tube on the probe.

Allow the zero air to flow for two minutes.


MAINTENANCE

AIR CAL


MAINTENANCE

WET

50


ppm

MAINTENANCE

CH

4

PUSH AIR