29 • Leak Testing

10,000 ppm Gas Tracer Operator’s Manual

on page 16. Select the desired display range. In the example below,
1000 ppm has been selected. Press and hold the (SHIFT)T button,
then press and hold the AIRS button and hold both until you hear
a beep. The Station ID Select Screen will display and the current
station ID will be flashing.

2. Use the (SHIFT)T and AIRS buttons to scroll to the desired station

ID, then press and release the POWER ENTER button to continue
to the Base Reading Screen. To return to the Station ID Select Screen
without taking a base reading, press and release the RESET
SILENCE button.

3. In order to get a base, or background, reading, take the instrument

a few feet away from where you intend to test. To save the base
reading, press and release the POWER ENTER button.