Replacing a sensor – RKI Instruments Gas Tracer 10,000 ppm User Manual

Page 99

Advertising
background image

10,000 ppm Gas Tracer Operator’s Manual

Replacing a Sensor • 92

4. Insert the battery pack into the charging station as shown in Figure

20 below so that the metal contacts on the back of the battery pack
come into contact with the metal contacts in the charging station.
When proper contact has been made, the red LED on the charging
station will turn on.

Figure 20: Putting the Battery Pack into the Charging Station

5. When a full charge has been reached, approximately 4 hours, the

red LED on the charging station will turn off. Remove the battery
pack from the charging station and unplug the charging station’s
power cord from the AC outlet.

Replacing a Sensor

1. Verify that the Gas Tracer is off.

2. Use a screwdriver to remove the three screws holding the belt clip

to the back side of the instrument. Two of these screws also secure
the flow chamber.

3. Remove the belt clip and unscrew the remaining screw holding in

the flow chamber.

4. Grab the exhaust tube and pull it gently away from the unit to slide

the flow chamber out of the unit.

5. Gently pull on the knob of the sensor gasket to remove it. Be sure

not to lose any of the filters that are in it. The sensors will be
exposed.

6. Remove the ppm sensor from its socket. All other sockets have a

Battery Pack

SDM-2012

BUMP

GX-2012 OFF

COPY

POWER

CAL.

EDIT

ENTER

1SEC ON
3SEC OFF