E - non-return to zero, B3 h – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 119

Advertising
background image

103

CHAPTER 7
Troubleshooting

E - Non-Return to Zero

Enter

Diagnostics

Return

to ZERO

Activate the

test and

review the

results.

Verify the
individual

load cell

signal outputs

PASS

PASS

Clean any material
from the scale

Check

mechanical

connections

are correct.

The zero tolerance

level set in

configuration has

been exceeded

Do not just

recalibrate or

increase the zero

tolerance levels.

Troubleshoot the

scale to why the

zero point has

shifted.

B3

H

Yes

E

No

The Return to Zero Test is used to determine
whether the instrument can still zero a scale
based on preset parameters. If you pass the
Return to Zero Test you are within the sum of the
preset Motion and Zero Tolerance settings. If you
fail you are outside the sum of the preset Motion
and Zero Tolerance settings. If you fail the test
there may be too much build up on the scale and
you need to clean the scale or you have scale
problems. You should do this test whenever you
cannot zero the scale.