CHAPTER 7
Troubleshooting

N - Blank Screen

N

No

Measure the excitation

voltage at J1

5 VDC?

Contact

Hardy Instrument

Technical Service

800-821-5831

Disconnect all the

connectors from the

back panel except

power.

Measure the

Excitation voltage

at J1

5 VDC?

1) Check for proper power at the
source connection.
2) Check the circuit breaker at the
source.

Check the power fuse located inside

the case

OK?

Replace the 2.5

amp slo-blow fuse
and supply power.

Does the fuse blow

out again?

1) Reconnect the jacks one
at a time checking the 5 VDC
excitation.
2) If reconnecting any jack
effects the 5 VDC, check for
wiring errors.

Contact Technical

Service

Monitor system for proper

operation

Check out complete

Measure AC Power

at J6

OK?

No

Yes

No

Yes

Yes

No

No

Display OK