Ok to fill input parameter, About ok to fill input parameter, Discharge parameters – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 48: About the discharge parameters, Ok to fill menu/setting ok to fill timer

Advertising
background image

HI-3010 Filler/Dispenser/IBC

32

Service Manual

OK to Fill Input Parameter

About OK to Fill Input Parameter

The OK to Fill Input Parameter is user selectable and is one
of the alarm settings. When the OK to Fill Input Parameter is
on, the instrument Auto Tares the scale and resets and starts
the Fill/Dispense timer. If for any reason the instrument can-
not detect the OK to Fill input before the OK to Fill timer
times out, the Not OK to Fill Alarm appears.

PARAMETER:OK TO FILL INPUT PARAMETER
RANGE: ON/OFF
DEFAULT: ON

FIG. 4-45 OK TO FILL INPUT PARAMETER

Step 1. Press the right or left arrow buttons to toggle

between ON and OFF.

Step 2. If you have turned the OK to Fill ON, press the

Enter button. The OK TO FILL MENU appears.
(See Fig. 4-46)

FIG. 4-46 OK TO FILL MENU/SETTING OK TO FILL

TIMER

PARAMETER:OK TO FILL TIMER
RANGE: 0-999
DEFAULT: 10 SECONDS

Step 3. Press the right or left arrow buttons to increase or

decrease the OK to Fill Time. In our example we set
the timer for 10 seconds.

Step 4. Press the Enter button to set the entry.
Step 5. Press the Exit button to return to the SETUP

MENU.

NOTE:

If you toggle the Ok to Fill to OFF, the OK to Fill
Menu will not appear.

Step 6. Press the down arrow until the cursor is in front of

Discharge. (See Fig. 4-47)

Discharge Parameters

About the Discharge Parameters

If your process requires you to discharge from the filled ves-
sel you need to turn on the Discharge function. If you turn
the Discharge ON, the Discharge Menu will display and you
can select between Auto Discharge, OK to Discharge, Proof
Switch Alarm, Switch Timer and/or the Aux (Auxiliary)
Device Timer. The Aux (Auxiliary) Device Timer turns on a
vibrator feeder or other device if the Not OK to Discharge
Alarm occurs. If you select OFF the Discharge Menu does
not display. Make sure you set the Zero Tolerance to deter-
mine when the fill vessel is empty (i.e. completed the dis-
charge).

PARAMETER:DISCHARGE
RANGE: ON/OFF
DEFAULT: OFF

PARAMETER:AUTO DISCHARGE
RANGE: ON/OFF
DEFAULT: OFF

PARAMETER:OK TO DISCHARGE
RANGE: ON/OFF
DEFAULT: OFF

PARAMETER:PROOF SWITCH
RANGE: ON/OFF
DEFAULT: OFF

PARAMETER:SWITCH TIMER
RANGE: 0-99
DEFAULT: 5 SECONDS

PARAMETER:AUX DEVICE TIMER
RANGE: 0-99
DEFAULT: 10 SECONDS

NOTE:

The description for these parameters is located in
the HI 3010 User Guide.

Step 1. To select Discharge press on the left or right arrow

buttons to toggle Discharge ON or OFF. (See Fig.
4-47)