CHAPTER 7
Troubleshooting

Step 9. Press the Exit button to return to the Choose Dis-

play.

Step 10. Press the Exit button to return to the Diagnostics

Display.

Stability Test

The Stability Test switches a fixed signal into the analog to
digital convertor, and calculates the mean squared variation
from the average reading, using 100 samples.
The test passes if the mean squared variation is less than 5.0,
and the average reading is between 30237 and 36955.

Step 1. Press the up or down arrow buttons until the cursor

is in front of Stability Test. (See Fig.7-76)

FIG. 7-76 DIAGNOSTICS DISPLAY/SELECTING

STABILITY TEST

Step 2. Press the Enter button. The Information display

appears “SYS. STABILITY TEST, PRESS ENTER
TO START”. (See Fig. 7-77)

FIG. 7-77 STABILITY TEST DISPLAY

Step 3. Press the ENTER button to perform the stability

test.

If the instrument passes the Stability Test
the Pass display appears. (See Fig. 7-78)
This means that the Mean Squared Varia-
tion is less than 5.0 and the average read-
ing is between 30237 and 36955. In short
the instrument is working fine.

FIG. 7-78 SYSTEM STABILITY TEST DISPLAY/

PASS

If the instrument does not pass the Stabil-
ity Test the Fail display appears. (See Fig.
7-79) This means that the Mean Squared
Variation is greater than 5.0 and/or the
average reading is not between 30237 and
36955. This test examines the internal
electronics and not the load cells input sig-
nal.

FIG. 7-79 SYSTEM STABILITY TEST DISPLAY/FAIL

NOTE:

In our example we show that the Means squared
variation is greater than 5.0 but the average
reading is within tolerance.

1.

Disconnect the power cord and recon-
nect the power cord to restart the
instrument.

2.

Repeat the Stability test.

3.

If the instrument Fails the Stability
Test again, contact Hardy Instruments
Inc., Technical Support for assistance.

Step 4. Press the Exit button to return to the Diagnostics

display.

Factory Defaults

CAUTION: I

F

YOU

CHOOSE

FACTORY

DEFAULTS

ALL

DATA

WILL

BE

LOST

! M

AKE

ABSOLUTELY

SURE

THAT

THIS

IS

WHAT

YOU

WANT

TO

DO

BEFORE

CHOOSING

THIS

OPTION

.

DO

NOT

USE

THIS

FUNCTION

IN

AN

EFFORT

TO

CORRECT

ANY

MALFUNCTIONS

IN

THE

OPERATION

OF

THE

INSTRUMENT

.

Step 1. Press the up or down arrow buttons until the cursor

is in front of Factory Defaults. (See Fig. 7-80)

Advertising