F - Verify Individual Load Cell Milli-Volt
Readings

F

Using a Multimeter

Physically lift the

signal leads to read
each load cell’s

output.

Check individual

load sensor

output mV

readings.

Check load sensor

Record load
sensor mV
output level for
comparison.

Repeat for all load

sensors

MV

readings

acceptable?

MV

readings

acceptable?

MV

readings

acceptable?

MV

readings

acceptable?

MV

readings

acceptable?

Using the load cell certificate, verify the milli volt per volt rating

Example: 3 mV/V load cells will produce approximately 15 mV at full load.

That is 5 volts excitation x 3 mV/V. A scale capacity of 1,000 lbs with 100 lbs

of dead load at empty the load point mV reading should equal 1.5 mV

1) No dead load.

Apply load and re-test.

2) Wiring error.

Verify color code using the
load cell certificate.
3) Open bridge circuit.
Disconnect power and verify

load point bridge resistance

reading with an Ohmmeter.

1) Stressed load cell, remove

all load and re-test.

2) Excessive loading.

For additional

testing go to:

H

K

A

Defective load cell?

Replace and repeat

Test F

If you were unable to determine the

Milli-volt readings. Go to

K load sharing

or

Contact Technical Support

TEST COMPLETE