Tare limit parameter, About the tare limit parameter, Setup menu/zero tolerance – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 56: Zero tolerance menu/zero tolerance, Zero tolerance menu/use auto zero, Zero tolerance menu/auto zero tolerance, Zero tolerance menu/auto zero time

Advertising
background image

HI-3010 Filler/Dispenser/IBC

40

Service Manual

DEFAULT: OFF

PARAMETER:AUTO ZERO TIME
RANGE: .01-9.99 SECONDS
DEFAULT: 1.00 SECOND

Step 1. Press the down button until the cursor is in front of

Zero Tolerance. (See Fig. 4-78)

FIG. 4-78 SETUP MENU/ZERO TOLERANCE

Step 2. Press the Enter button. The Zero Tolerance Menu

appears with the cursor in front of Zero Tolerance.
(See Fig. 4-79)

FIG. 4-79 ZERO TOLERANCE MENU/ZERO

TOLERANCE

Step 3. Press the Clear button to clear the current value.
Step 4. Used the alphanumeric key pad to enter the new

Zero Tolerance value.

Step 5. Press the Enter button to set the entry.

NOTE:

Zero Tolerance should be set if you are going to
use the Discharge Function. Setting the zero tol-
erance sets the parameter that indicates when a
vessel is empty after discharging a fill.

Step 6. Press the down arrow button until the cursor is in

front of Use Auto Zero. (See Fig. 4-80)

FIG. 4-80 ZERO TOLERANCE MENU/USE AUTO

ZERO

Step 7. Press the right or left arrow buttons to toggle

between OFF or ON.

Step 8. Press the Enter button to set the entry.
Step 9. Press the down arrow button until the cursor is in

front of AZERO Tol. (See Fig. 4-81)

FIG. 4-81 ZERO TOLERANCE MENU/AUTO ZERO

TOLERANCE

Step 10. Press the Clear button to clear the current value.
Step 11. Use the alphanumeric key pad to enter the new

Auto Zero Tolerance value.

Step 12. Press the Enter button to set the entry.
Step 13. Press the down arrow button until the cursor is in

front of Auto Zero Time. (See Fig. 4-82)

FIG. 4-82 ZERO TOLERANCE MENU/AUTO ZERO

TIME

Step 14. Press the Clear button to clear the current value.
Step 15. Use the alphanumeric key pad to enter the new

time.

Step 16. Press the Enter button to set the entry.
Step 17. Press the Exit button to return to the SETUP

MENU.

Tare Limit Parameter

About the Tare Limit Parameter

The Tare Limit Parameter limits the amount of automatic
tare. Tare is the artificial zeroing of the weight hopper so that
a new weight can be displayed. Also, the action of adjusting
out the know weight of the container from the total indicated
weight, so that the indicator reads net weight directly.

PARAMETER:TARE LIMIT
RANGE: .000001-999999
DEFAULT: 999999

Step 1. Press the down arrow until the cursor is in front of

Tare Limit. (See Fig. 4-83)