Voltage-controlled current apg setup – AMETEK XG Family Programmable DC User Manual

Page 112

Advertising
background image



Analog

Programming (APG) and Isolated Analog Programming (ISOL)

4-12

M370430-01 Rev E

Voltage-Controlled Current APG Setup

Activating APG Current Mode

To activate APG current mode using an external voltage source:

1.

Turn the 9-position mode control to the CAP position or press
the rotary Adjust/Enter control if the mode control is already
at the CAP position.

is displayed on the output voltage display.

2.

Turn the rotary Adjust/Enter control to select source

(Voltage source from non-isolated connector).

3.

Press the rotary Adjust/Enter control to commit the selected

mode.

is displayed on the output voltage display.

4.

Select the input source level using the rotary Adjust/Enter

control.

5.

Once the desired level has been selected, press the rotary

Adjust/ Enter control to commit the change.

The display will return to the normal display.

Deactivating APG Current Mode

To deactivate the APG current mode:

1.

Turn the 9-position mode control to the CAP position or press
the rotary Adjust/Enter control if the mode control is already
at the CAP position.

2.

ΧΑΠρ

is displayed on the output voltage display.

3.

Turn the rotary Adjust/Enter control to select

.

4.

Press the rotary Adjust/Enter control to commit the change.

The display will return to the normal display.