Calibration

and Troubleshooting

7-2

M370430-01 Rev E

Introduction

The calibration of the power supply is software dependent, and
there are no potentiometers to adjust.

Calibration is performed via SCPI commands. The following items
need to be calibrated:

Programmed voltage

Voltage readback

Programmed current

Current readback

Non-isolated voltage monitoring

Isolated voltage monitoring

Non-isolated current monitoring

Isolated current monitoring

Isolated source for analog programming

Non-isolated source for analog programming

Over voltage protection system

CAUTION: Equipment damage

Calibration procedures should only be performed by qualified user.