AMETEK XG Family Programmable DC User Manual

Page 285

Advertising
background image



Calibration and

Troubleshooting

M370430-01 Rev E

7-21

7

Isolated Resistive Programming of Current Calibration

To calibrate the isolated resistive programming of current:

1.

Short the main output with a shunt. Connect a multimeter to
measure the voltage across the shunt.

2.

Connect the 4.000 kΩ resistor to the isolated connector. One

terminal to resistive programming of current (J3.7), and
voltage programming of current (J3.3), and the other to
common (J3.2).

3.

Set the Resistive APG level to 4 kΩ The SCPI command is:

SYST:COMM:APR:LEV:CURR 4

4.

Set the input source as Resistive Isolated for Current

Programming. The SCPI command is:

SYST:REM:SOUR:CURR IAR

5.

Turn the main output On or send the command:

:OUTP ON

6.

Enter the interactive calibration mode by sending the SCPI

command:

CAL:OUTP:ANAL:CURR:RES:ISOL

7.

Press the terminal keys - or + to reduce or increase the main

output current value. Adjust the main output current to 100%.

8.

Press any key other than the + or - keys to exit from the

interactive calibration mode.