Calibration and

Troubleshooting

M370430-01 Rev E

7-13

7

Non-isolated Resistive Programming of Voltage Calibration

Important:

For maximum accuracy at a specific APG level, follow the

calibration procedure again but use the intended operating resistance level in
steps 2 and 3.

To calibrate the non-isolated resistive programming of voltage:

1.

Disconnect the load and connect the voltmeter to the output.

2.

Connect the 4.000kΩ resistor to the non-isolated connector,

resistive programming of voltage, and voltage programming
(J1.9-J1-11, and J1.7).

3.

Set the resistive APG level to 4kΩ The SCPI command is:

SYST:COMM:APR:LEV 4

4.

Set the input source as Resistive Non-Isolated for Voltage

Programming. The SCPI command is:

SYST:REM:SOUR ARES

5.

Turn the main output On by pressing the OUTPUT ENABLE

Main button on the front panel or send the command:

:OUTP ON

The OUTPUT ENABLE Main button will illuminate.

6.

Enter the interactive calibration mode by sending the SCPI

command:

CAL :OUTP:ANAL:RES

7.

Press the terminal keys - or + to reduce or increase main

output voltage value. Adjust the main output voltage to
100%.

8.

Press any key other than the + or - keys to exit from

interactive calibration mode.