Calibration

and Troubleshooting

7-14

M370430-01 Rev E

Non-isolated Voltage Programming of Current Calibration

Important:

For maximum accuracy at a specific APG level, follow the

calibration procedure again but use the intended operating voltage level in steps
3 and 4.

To calibrate the non-isolated voltage programming of current:

1.

Short the main output with a shunt. Connect a multimeter to
measure the voltage across the shunt.

2.

Connect the input voltage source to the non-isolated

connector, current programming (J1.10 - J1.7).

3.

Set the Voltage APG level to 4 V. The SCPI command is:

SYST:COMM:APR:LEV:CURR 4

4.

Set the input source voltage to 4.000 V.

5.

Set the input source as Voltage Non-Isolated for Current. The

SCPI command is:

SYST:REM:SOUR:CURR AVOL

6.

Turn the main output On by pressing the OUTPUT ENABLE

Main button on the front panel or send the command:

:OUTP ON

The OUTPUT ENABLE Main button will illuminate.

7.

Enter the interactive calibration mode by sending the SCPI

command:

CAL:OUT P:ANAL:CURR

8.

Press the terminal keys - or + to reduce or increase the main

output current value. Adjust the main output current to 100%.

9.

Press any key other than the + or - keys to exit from

interactive calibration mode.