Active Range Thresholds

NOTE: These settings apply to the Detection Range selected, and override the
Global Thresholds specified on the Basic Settings tab (described on page 3-19)
.

%BP Intensity

See “%Base Peak Intensity” on page 3-20.

%Max Peak Area

See “%Max Peak Area” on page 3-20.

Minimum Intensity

Specifies the absolute peak intensity below which peaks are
not detected. Calculated relative to zero.

NOTE: Maximum possible intensity for Mariner data is
11,000 counts (spectrum that has not been accumulated or
summed), and for Voyager data is 66,000 counts.

NOTE: This parameter was previously named Absolute
Threshold.

Minimum Area

Specifies the peak area below which peaks are not detected.
Calculated relative to peak valleys.

Hint: Display the peak list to determine the appropriate area
value to enter.

Noise Threshold

Used to determine peak boundaries. Click the field, then
right-click-drag over a segment of the trace to automatically
set.

For more information, see “Process that Occurs During Peak
Detection, Centroiding, and Integration” on page 3-67.

Continued

Table 3-4 Advanced Settings (Spectrum Data Only) (Continued)