Manual Calibration

Data Explorer

Software User’s Guide

5-5

5

5.3 Manual Calibration

This section describes:

Overview of manual calibration

Manually calibrating

Creating or modifying a calibration reference file

Reverting to instrument calibration

Hints for calibrating Mariner data

Hints for calibrating Voyager data

NOTE: Manual calibration is not supported for Mariner
DAD data.

5.3.1 Overview of Manual Calibration

Overview

During manual calibration:

You specify a calibration reference file (.REF) that
contains reference masses.

The software matches peaks by comparing observed
masses in the spectrum to reference masses.

The software lists masses that match within the specified
peak matching criteria, and calculates calibration
constants.

You can apply the calibration constants to all spectra in a
Mariner or Voyager data file.

You can apply individual calibration constants to each
spectrum in a Voyager data file.